Direct Determination of Chromium in Blood Serum by Electrothermal-Atomization Atomic Absorption Spectrometry

Autoři

  • O. Mestek Department of Analytical Chemistry, Institute of Chemical Technology, Prague,
  • T. Zima First Institute of Medical Chemistry and Biochemistry, First Medical Faculty, Charles University, Prague
  • M. Suchanek Department of Analytical Chemistry, Institute of Chemical Technology, Prague,
  • J. Zilkova Department of Analytical Chemistry, Institute of Chemical Technology, Prague,

Abstrakt

The method of direct determination of Cr by ETAAS in blood serum and whole blood after dilution with 0.1% Triton X-100 and 1% HNO3 is described. The temperature programme of graphite furnace involved careful drying at 70 oC, two-step pyrolysis at 800 and 1100 oC, and atomization at 2400 oC. No chemical modification was included, background correction was carried out by deuterium arc corrector. The detection limit was 0.24 mg.l-1 Cr. The method was used for analysis of blood serum of patients with chronic renal failure treated by heamodialysis.

Publikováno

15.10.1998

Jak citovat

Mestek, O., Zima, T., Suchanek, M., & Zilkova, J. (1998). Direct Determination of Chromium in Blood Serum by Electrothermal-Atomization Atomic Absorption Spectrometry. Chemické Listy, 92(9). Získáno z http://www-.chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/2674

Číslo

Sekce

Články